Specification

NaI-63×63 Scintillation Detector Assembly includes Ø63.5×63.5 mm  (Ø2.5”x2.5”) NaI(Tl) crystal coupled to Hamamatsu R6232 PMT hermetically packed in an Aluminum housing with 0.5 mm thick walls and integrated magnetic shield and temperature sensor (optional)

Detector type NaI(Tl)
Detector size Ø63.5×63.5 mm
Energy range From 10 keV
Typical resolution 6.5% at 662 keV
Ambient dose rate N/A
Typical sensitivity,
cps/(µSv·h-1)
2300 (137Cs)
1970 (60Co)
Neutron detection No
Readout  PMT+JEDEC B14-38
Dimensions Ø70×198 mm
Weight 900 g